In CT imaging, the window level (centre) is set to +40 HU for a soft-tissue window. What is the primary effect of changing the window level to +400 HU while keeping the window width constant?
- A The spatial resolution improves
- B The image appears overall brighter
- C Image contrast is lost completely
- D The image appears overall darker ✓
Correct answer: D. The image appears overall darker
Explanation
The window level sets the centre of the grey scale. D higher window level means that higher attenuation values are mapped to mid-grey, so soft tissue (around +40 HU) falls below the centre and appears darker. Window level controls overall brightness on the displayed image.
Reference: Bushberg, The Essential Physics of Medical Imaging, 3rd ed.