Radiology · Fundamentals of X-Ray, CT, MRI and USG (Physics, Basics)

In CT, a ring artifact on the reconstructed image is most commonly caused by:

  • A Patient motion during the scan
  • B A miscalibrated or defective detector element
  • C Beam hardening from dense bone
  • D Insufficient tube current (mAs)
Correct answer: B. A miscalibrated or defective detector element

Explanation

Ring artifacts arise from a faulty or miscalibrated detector channel that consistently reads incorrect attenuation values. As the tube rotates, this detector traces a ring centered on the isocenter. Patient motion causes blurring or streaking, beam hardening causes dark bands between dense objects, and low mAs causes quantum mottle (grainy noise), not rings.

Reference: Bushberg, The Essential Physics of Medical Imaging, 3rd ed.