In CT, a ring artifact on the reconstructed image is most commonly caused by:
- A Patient motion during the scan
- B A miscalibrated or defective detector element ✓
- C Beam hardening from dense bone
- D Insufficient tube current (mAs)
Correct answer: B. A miscalibrated or defective detector element
Explanation
Ring artifacts arise from a faulty or miscalibrated detector channel that consistently reads incorrect attenuation values. As the tube rotates, this detector traces a ring centered on the isocenter. Patient motion causes blurring or streaking, beam hardening causes dark bands between dense objects, and low mAs causes quantum mottle (grainy noise), not rings.
Reference: Bushberg, The Essential Physics of Medical Imaging, 3rd ed.