Orthopedics · Joint Replacement — Advanced (THR/TKR Complications, Revision, Bearings, Periprosthetic Fractures)

A 62-year-old man with a well-functioning 36 mm metal-on-highly-crosslinked-polyethylene total hip replacement presents with groin pain at 7 years. Serum cobalt is elevated although there is no significant polyethylene wear on radiographs. CT shows a small fluid collection at the joint. Which mechanism best explains these findings?

  • A Wear-through of the polyethylene liner exposing the metal shell
  • B Fretting and crevice corrosion at the modular head-neck trunnion junction
  • C Corrosion of the titanium acetabular shell screws
  • D Oxidative degradation of the polyethylene producing inflammatory debris
Correct answer: B. Fretting and crevice corrosion at the modular head-neck trunnion junction

Explanation

Elevated cobalt with minimal polyethylene wear points to trunnionosis, mechanically assisted crevice corrosion and fretting at the modular head-neck junction. Risk factors include large femoral heads, mixed-alloy couples, male sex, high body mass index, and high offset. It can produce an adverse local tissue reaction identical to that seen after metal-on-metal arthroplasty. Liner wear-through would show eccentric polyethylene thinning, and screw corrosion would not selectively raise cobalt over titanium.

Reference: Miller's Review of Orthopaedics, 8th ed.

High-yield for: NEET PGINI-CETNExTFMGEUSMLEPLABMRCP

Written and medically reviewed by the StethoPrep medical team.

Sponsored

Want to test yourself?

Create a free account for timed mock tests, mistake tracking, and FSRS spaced-repetition revision across 43,000+ MCQs.

Start free → Log in

More Joint Replacement — Advanced (THR/TKR Complications, Revision, Bearings, Periprosthetic Fractures) MCQs

See all Joint Replacement — Advanced (THR/TKR Complications, Revision, Bearings, Periprosthetic Fractures) MCQs →