A 62-year-old man with a well-functioning 36 mm metal-on-highly-crosslinked-polyethylene total hip replacement presents with groin pain at 7 years. Serum cobalt is elevated although there is no significant polyethylene wear on radiographs. CT shows a small fluid collection at the joint. Which mechanism best explains these findings?
- A Wear-through of the polyethylene liner exposing the metal shell
- B Fretting and crevice corrosion at the modular head-neck trunnion junction ✓
- C Corrosion of the titanium acetabular shell screws
- D Oxidative degradation of the polyethylene producing inflammatory debris
Explanation
Elevated cobalt with minimal polyethylene wear points to trunnionosis, mechanically assisted crevice corrosion and fretting at the modular head-neck junction. Risk factors include large femoral heads, mixed-alloy couples, male sex, high body mass index, and high offset. It can produce an adverse local tissue reaction identical to that seen after metal-on-metal arthroplasty. Liner wear-through would show eccentric polyethylene thinning, and screw corrosion would not selectively raise cobalt over titanium.
Reference: Miller's Review of Orthopaedics, 8th ed.