A craniotomy set was opened but unused during surgery on a patient later found to have sporadic Creutzfeldt-Jakob disease. The CSSD must decide how to manage the instruments that had contact with the operative field. Which reprocessing recommendation is correct for these potentially prion-contaminated reusable instruments?
- A Decontamination with 1N sodium hydroxide for 60 minutes followed by prolonged prevacuum steam sterilization at 134 degrees Celsius, or disposal of the instruments if possible ✓
- B Ethylene oxide sterilization, because prions are destroyed by alkylating agents
- C Routine cleaning followed by standard steam sterilization at 134 degrees Celsius for 3 minutes
- D Immersion in 2% glutaraldehyde for 20 minutes followed by normal rinsing
Explanation
Prions resist ordinary steam cycles, ethylene oxide, formaldehyde and glutaraldehyde. Guideline recommendations for high-risk tissue contact are destruction of the instrument, or decontamination in 1N sodium hydroxide followed by an extended steam cycle at 134 degrees Celsius. Option C fails because the short 3-minute cycle does not inactivate prions, and option D is merely high-level disinfection, inadequate even against spores. Single-use neurosurgical instruments are preferred in such cases.
Reference: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008 ed.