Anaesthesia · Monitoring in Anaesthesia (CNS, CVS, Respiratory)

A patient receiving prolonged nitric oxide therapy for acute respiratory distress syndrome develops increasing methaemoglobinaemia. The conventional two-wavelength pulse oximeter will display a saturation closest to:

  • A Rising toward 100%
  • B Zero, because MetHb absorbs equally at both wavelengths
  • C Accurate values until MetHb exceeds 20%
  • D Approximately 85%, regardless of the actual arterial saturation
Correct answer: D. Approximately 85%, regardless of the actual arterial saturation

Explanation

Methaemoglobin absorbs light almost equally at both the 660 nm and 940 nm wavelengths used by standard pulse oximeters, causing the device to read a fixed saturation of approximately 85% irrespective of true arterial oxygenation. This occurs because MetHb has similar absorbance ratios to a mixture containing roughly 85% oxyhaemoglobin. Co-oximetry measuring multiple wavelengths is required for accurate assessment.

Reference: Miller's Anesthesia, 9th ed.

High-yield for: NEET PGINI-CETNExTFMGEUSMLEPLABMRCP

Written and medically reviewed by the StethoPrep medical team.