An anaesthetist working at a high-altitude centre where barometric pressure is 630 mmHg maintains a patient on sevoflurane. Compared with sea level practice, what adjustment is needed to achieve the same anaesthetic depth?
- A The vaporiser dial setting must be reduced because alveolar uptake is faster at altitude
- B No adjustment is needed because MAC is expressed as a universal percentage
- C The vaporiser dial setting must be increased because anaesthetic depth depends on partial pressure, not percentage concentration ✓
- D The vaporiser must be recalibrated to deliver a fixed partial pressure irrespective of dial setting
Explanation
Anaesthetic depth depends on the partial pressure of the agent in the brain, not the fractional concentration. At 630 mmHg barometric pressure, 2 percent sevoflurane delivers only about 12.6 mmHg versus 15.2 mmHg at sea level, so a higher dial setting, roughly 2.4 percent, is required to reach one MAC. Variable-bypass vaporisers output a percentage concentration, so the clinician must compensate manually for reduced ambient pressure.
Reference: Miller's Anesthesia, 9th ed.